Acculab LA-110, Acculab LT-3200 has been tested in Datalyse
Menu in Datalyse
Graphical measuring, there is only one measuring method, the fast one.
Baud rate 1200, odd parity, 1 stop, 7 data bits.
No version command. The cable is device specific and not included
Data Interface Cable PIN Configuration For VIR-Series Balances. See www.acculab.com.
Acculab VIR-Series | Standard PC RS-232 |
Balance 9-PIN port | 9-PIN Connector |
RxD 2 | 3 TxD |
TxD 3 | 2 RxD |
Signal Ground 5 | 5 Signal Ground |
Acculab VIR-Series | Standard PC RS-232 |
Balance 9-PIN port | 25-PIN Connector |
RxD 2 | 2 TxD |
TxD 3 | 3 RxD |
Signal Ground 5 | 7 Signal Ground |
Acculab is made by Sartorius.